Manipal Institute of Technology
Ion Chromatographic System (IC)
Central Instrumentation Facility Academic Block – 4, Ground Floor, Innovation Centre, MIT, Manipal, 576104, India
Description
Technique used to separate charged particles from a liquid and measure their concentration. IC systems can analyze particles such as anions, cations, organic salts, and proteins.
Specifications
- Maximum Operating Pressure: 35MPa (5000psi)
- Flow Rate Range: 0.05-5.0 mL/min in 0.01 increment
- Flow Precision: <0.2%
- Flow Accuracy: <1.0% from 0.4 to 2.0mL/min at 1000
- Resolution of conductivity detector: 0.1 nS
- Sample holders : 50 polyvials of 5mL size
- Air conditioned (21 to 24 degree C)
- Nitrogen with 99.99% purity, regulator (0-1 bar)
Limitations
- Waste and chemically treated water/solution
Testing Capabilities
- Major cations like Na,K, Ca, Mg can be identified
- Major anions like F, Cl, Pho can be identified
Sample Requirements
- Water samples need to be filtered (0.20µm)
- Conductivity should be within 120µS
Types of samples
- Water Sample – Ground, Surface, Rainwater etc
Details
Brand: Dionex ICS-1100
Model: N/A
Rating
-- / 5 (No reviews yet)
Admins
CIF
Book this Instrument
To make a booking, fill out the booking form.
Base Price
₹845
Addon
₹0
Total
₹845